AS7030B-EVALKIT - PPG/ECG Biometric Sensor Eval Kit | ams OSRAM
부품 번호: AS7030B-EVALKIT ✓ 활성| 수량 | 단가 | 합계 |
|---|---|---|
| 1 | $0 | $0.00 |
| 10 | $0 | $0.00 |
| 100 | $0 | $0.00 |
| 500 | $0 | $0.00 |
| 1,000 | $0 | $0.00 |
AS7030B-EVALKIT 개요
A biometric sensor evaluation kit is a complete hardware and software development platform that lets engineers characterize a sensor before committing it to a wearable or medical product design. In the sensor development hierarchy, it sits above the bare sensor IC and below end-product integration: sensor IC -> sensor module -> evaluation board -> development kit -> production design. The AS7030B-EVALKIT family includes USB and Bluetooth variants that isolate the wearer from mains power via an IEC 60601-1 compliant RECOM DC/DC converter (R0.25S-0505/H or R0.25S-0505/HP) and signal isolators.
Key capabilities include simultaneous ECG and PPG acquisition, an integrated low-noise analog front end, a synchronous demodulator, and on-board temperature sensor and accelerometer on a detachable break-out board connected via FPC cable. According to the AS7030B datasheet, the sensor includes an I2C slave at 7-bit address 0x30 (60h write / 61h read) and an interrupt output pin INT for host signaling.
Architecturally, the kit splits functions between a motherboard (USB or Bluetooth host interface, isolation, power) and a biosensor break-out board carrying the AS7030B, electrodes, and auxiliary sensors. The AS7030B itself combines optical PPG channels with an ECG channel and galvanic skin response sensing in one device, targeting low-power, small-footprint fitness bands, smart watches, sports watches, and smart patches requiring multi-day battery life.
Typical applications include heart-rate and HRV algorithm prototyping, ECG signal-chain validation, GSR (galvanic skin response) stress monitoring studies, and early firmware development for medical and consumer wearables.
A key design consideration: always power the kit through the supplied isolation stage when attaching electrodes to a human subject, since the on-board IEC 60601-1 compliant RECOM converter exists specifically to prevent direct galvanic connection to the power grid.
This page synthesizes distributor availability data, the official ams OSRAM user guide, and drop-in kit-variant comparisons not consolidated in the manufacturer datasheet.
드롭인 대체품: AS7030B-EVALKIT — 동일 패키지, 핀 호환. PCB 재작업이 필요한 다른 패키지 부품은 제외됩니다.
빠른 비교 도구 — 나란히 비교할 대체 부품을 선택하십시오:
AS7030B_EVALKIT_USB
✅ 드롭인📋 참조 대체품 (카탈로그에 없음)
AS7030B_EVALKIT_BT
✅ 드롭인📋 참조 대체품 (카탈로그에 없음)
AS703x_EVM_SN
✅ 드롭인📋 참조 대체품 (카탈로그에 없음)
AS7038GB-EVALKIT
✅ 드롭인📋 참조 대체품 (카탈로그에 없음)
AS7038RB-EVALKIT
✅ 드롭인📋 참조 대체품 (카탈로그에 없음)
AS7030B-EVALKIT 사양 (제조사 공개)
| Product Type | Biometric Sensor Evaluation Kit |
| Target Sensor | AS7030B biosensor (ECG, PPG, GSR) |
| Sensed Parameters | ECG, photoplethysmography (PPG), galvanic skin response (GSR) |
| Communication Interface | I2C slave (7-bit address 0x30, 60h write / 61h read) |
| Interrupt Output | INT pin for host interrupt |
| Host Connection | USB or Bluetooth (variant dependent) |
| Supported OS | Windows |
| Isolation | IEC 60601-1 compliant RECOM DC/DC converter R0.25S-0505/H or R0.25S-0505/HP |
| On-Board Sensors | Temperature sensor, accelerometer, electrodes on break-out board |
| Board Architecture | Motherboard + detachable biosensor break-out board via FPC cable |
| Auxiliary Sensor Option | Supports AS7038GB and AS7038RB biosensor variants |
| Analog Front End | Low-noise AFE with synchronous demodulator |
| Typical Applications | Fitness bands, smart watches, sports watches, smart patches |
AS7030B-EVALKIT 인터페이스 및 커넥터
AS7030B-EVALKIT에 대해 제조사에서 공개한 인터페이스 목록이 없습니다. 커넥터 및 헤더 세부 정보는 제조사 문서를 참조하십시오.
다음을 참조하십시오: 제조사 문서 전체 확장 헤더 핀아웃 및 커넥터 핀 번호는 다음을 참조하십시오.
일반적인 애플리케이션
AS7030B-EVALKIT 적합 용도: 6 애플리케이션s: Fitness Band and Smart Watch Development, ECG Signal Chain Validation, GSR Stress and Hydration Monitoring Research, Smart Patch and Multi-Day Battery Wearables, Medical Wearable Prototyping and Algorithm Development, Optical Sensor Education and Lab Training.
Fitness Band and Smart Watch Development
The AS7030B-EVALKIT directly supports wearable heart-rate product development because the AS7030B's low-power design and small form factor are explicitly targeted at fitness bands, smart watches, and sports watches per the ams OSRAM datasheet. Engineers use the kit's detachable break-out board to test optical PPG placement on a wrist or chest form factor while streaming data over USB or Bluetooth to the PC GUI. The synchronous demodulator in the low-noise analog front end suppresses ambient light interference, and the INT interrupt pin maps naturally to motion-triggered sampling in firmware. Validated register configurations transfer directly from kit evaluation to production PCB, reducing design risk before layout commitment.
추천
ECG Signal Chain Validation
For ECG algorithm and analog-front-end validation, the kit provides an embedded ECG channel with a low-noise AFE, electrodes on the break-out board, and a safety isolation stage using the IEC 60601-1 compliant RECOM R0.25S-0505/H DC/DC converter. According to the ams OSRAM user guide, isolator ICs on all signals guarantee no physical connection between subject electrodes and the power grid, enabling safe on-body capture over USB power. Engineers can validate lead-off handling, noise floor, and sampling behavior in the PC GUI before porting the register map to a production microcontroller such as an XMC series MCU over the documented I2C address 0x30.
추천
GSR Stress and Hydration Monitoring Research
The AS7030B integrates galvanic skin response (GSR) sensing, and the evaluation kit exposes this channel alongside ECG and PPG so researchers can study stress, hydration, and electrodermal activity with a single instrument. MIKROE's ECG/GSR Click confirms the AS7030B is positioned by the manufacturer as a vital-sign sensor covering both ECG and GSR operations. Using the kit's PC software, investigators log simultaneous GSR and heart-rate waveforms during controlled stimulus protocols, using the temperature sensor and accelerometer on the break-out board as contextual channels. This multi-modal capture shortens study cycles compared with discrete GSR plus PPG instrumentation, while the USB isolation stage maintains subject safety.
추천
Smart Patch and Multi-Day Battery Wearables
Smart patches impose strict energy budgets, and the ams OSRAM datasheet states the AS7030B's low-power design suits applications where users expect extended, multi-day intervals between battery recharges. The evaluation kit lets designers quantify this claim empirically: by configuring LED duty cycles, ECG sampling rates, and interrupt-driven operation through the INT pin, firmware teams measure actual average current on the kit before committing a battery. The detachable break-out board emulates a patch's flexible substrate placement, and the Bluetooth variant enables realistic motion testing untethered from the bench. Results feed directly into battery sizing and duty-cycle decisions for the final patch product.
추천
Medical Wearable Prototyping and Algorithm Development
Medical device teams use the AS7030B-EVALKIT to prototype vital-sign algorithms ahead of formal design controls. The kit delivers synchronized ECG and PPG streams plus contextual temperature and motion data through a documented interface (I2C address 0x30, INT interrupt), letting algorithm engineers develop heart-rate variability, pulse transit time, or arrhythmia-detection prototypes against real captured data. The IEC 60601-1 compliant isolation, implemented with the RECOM R0.25S-0505/H converter and signal isolators, means subject recordings can be gathered under essentially the same safety architecture that a production device will require, per the ams OSRAM user guide. This reduces the gap between feasibility studies and design-verification testing.
추천
Optical Sensor Education and Lab Training
Universities and training labs adopt the AS7030B-EVALKIT as a hands-on platform for biomedical instrumentation courses because it exposes a complete photoplethysmography, ECG, and GSR signal chain with professional isolation practice built in. Students observe the synchronous demodulator's effect on motion artifact, configure the I2C register map at address 0x30, and capture their own physiological signals safely thanks to the IEC 60601-1 compliant RECOM isolation converter documented in the user guide. The AS7030B USB EVALUATION KIT is listed by DigiKey as an optical sensor evaluation board and ships same-day when stocked, making it practical for semester lab setups. The Windows GUI requires no embedded programming for first experiments.
추천
추천 제품 요약
엔지니어링 참조 데이터: AS7030B-EVALKIT — 비교, 설계 지침 및 규정 준수 정보.
선택 가이드
대체품과의 비교
| 파라미터 | 이 제품 | AS7030B_EVALKIT_USB | AS7030B_EVALKIT_BT | AS703x_EVM_SN |
|---|---|---|---|---|
| Package | Evaluation Kit (motherboard + break-out board) | Evaluation Kit (motherboard + break-out board) - same | Evaluation Kit (motherboard + break-out board) - same | Evaluation Kit (motherboard + break-out board) - same |
| Brand | ams OSRAM | ams OSRAM | ams OSRAM | ams OSRAM |
| Target Sensor | AS7030B | AS7030B | AS7030B | AS7030B / AS7038GB / AS7038RB |
| Measured Signals | ECG, PPG, GSR | ECG, PPG, GSR | ECG, PPG, GSR | ECG, PPG (per sensor variant) |
| Detach Break-Out Board | Yes (FPC cable) | Yes (FPC cable) | Yes (FPC cable) | Yes (FPC cable) |
| I2C Sensor Address | 0x30 (7-bit) | 0x30 (7-bit) | 0x30 (7-bit) | 0x30 (7-bit, AS7030B) |
주요 차별점
- Multi-modal sensing on one platform (vs AS7030B_EVALKIT_USB)
- Untethered evaluation option (vs AS7030B_EVALKIT_USB)
- Broader sensor coverage in one kit family (vs AS703x_EVM_SN)
- Patient-safety isolation built in (vs AS7038GB-EVALKIT)
설계 노트
Never bypass or remove the on-board isolation when attaching electrodes. The kit deliberately includes an IEC 60601-1 compliant RECOM DC/DC converter (R0.25S-0505/H or R0.25S-0505/HP) plus isolator ICs on all signals so that, per the ams OSRAM user guide, there is no physical connection between the electrodes and the power grid when USB-powered. Powering the motherboard from an external supply or substituting a non-isolated converter defeats this protection and creates a patient-safety hazard.
Use the FPC cable between motherboard and biosensor break-out board to explore real mechanical placement before finalizing a wearable PCB. The manufacturer deliberately placed the AS7030B, temperature sensor, accelerometer, and electrodes on the detachable board so it can be used in any application. During PPG evaluation, shield the optical window from ambient light; the synchronous demodulator rejects interference best when LED current and sampling settings in the GUI are tuned together rather than independently.
When porting from the kit to a production board, respect the AS7030B's I2C timing at 7-bit address 0x30 (60h write / 61h read per datasheet PD000544) and use the INT interrupt output rather than polling to reduce bus traffic and average power. Keep the analog front end's supply clean: the kit's low-noise AFE performance depends on the isolated 5 V rail, so replicate the local decoupling shown on the evaluation board in your own layout.
규정 준수 정보
The evaluation kit integrates an IEC 60601-1 compliant RECOM DC/DC converter (R0.25S-0505/H or R0.25S-0505/HP) for patient isolation, per the ams OSRAM user guide. Kit-level RoHS/REACH status not stated in verified data.