AD515AKH - 150 fA FET-Input Electrometer Op Amp | Analog Devices
N.º da peça: AD515AKH ✓ AtivoAD515AKH Visão geral
An electrometer operational amplifier is a low-noise, high-input-impedance amplifier optimized for measuring very small currents and voltages without loading the source. In the component hierarchy, the AD515AKH is a precision operational amplifier, an analog amplifier IC, and a member of the broader analog and mixed-signal semiconductor category. Its FET input stage enables measurement systems to preserve signal integrity when connected to high-impedance transducers, photodiodes, chemical sensors, and other sources with source impedances in the gigaohm range.
Key features include a guaranteed maximum input bias current of 150 fA, low-power operation with a 0.6 mA typical quiescent current, and a maximum quiescent current of 1.5 mA. The AD515A family also provides laser-trimmed offset voltage, low offset drift, and low noise. These characteristics allow the AD515AKH to support stable dc measurements while avoiding excessive signal-source loading. Its precision architecture is particularly useful when measurement accuracy is governed more by input bias current and offset behavior than by high bandwidth.
The AD515AKH represents a monolithic successor to the original AD515 electrometer and can replace that earlier device in most applications. Compared with general-purpose bipolar-input operational amplifiers, the AD515A topology uses a FET input stage to reduce input bias current by many orders of magnitude. The K-grade AD515AKH is specified for a maximum bias current of 150 fA, while the L-grade AD515AL reaches 75 fA and the J-grade AD515AJ reaches 300 fA. This grading lets designers balance leakage limits, availability, and application requirements without changing the basic measurement architecture.
Typical uses include pH and ion-selective electrode interfaces, photodiode and optical-sensor amplifiers, high-impedance voltage dividers, capacitance sensing, laboratory instrumentation, battery and gas monitoring, and precision current-to-voltage converters. In each case, the amplifier is placed directly at the high-impedance input so that cable and board leakage, guarding, clean power, and stable reference grounding become as important as closed-loop resistor accuracy.
A critical design practice is to surround sensitive inputs with guarding and keep input traces short and clean. Although input bias current is limited to 150 fA, board contamination, moisture, flux residue, and adjacent high-impedance nodes can increase leakage substantially. Use a shielded or guarded PCB layout, clean conformal coating where environmental exposure is expected, and low-value feedback resistors when system bandwidth permits.
This page combines the manufacturer product description, verified distributor pricing, package details, cross-reference findings, and practical design guidance for engineers evaluating the AD515AKH. It also distinguishes pin-compatible family options from broader functional equivalents so that replacement decisions are based on electrical proximity rather than package similarity alone.
Alternativas drop-in para AD515AKH — mesmo encapsulamento, compatível pino a pino. Peças de encapsulamento diferente que exigem retrabalho na PCB são excluídas.
Variantes desta série
Modelos da mesma série compatíveis como drop-in com AD515AKH (mesmo formato e footprint) — diferindo em Product Type, Amplifier Type, Input Offset Voltage, Maximum, Input Transistor Type.
AD515ALH
✅ Drop-In ⚠️ Especificações não verificadas📋 Alternativa de referência (fora do catálogo)
AD515AJH
✅ Drop-In📋 Alternativa de referência (fora do catálogo)
AD517SH
✅ Drop-InConsulte o preço
Ver ficha técnica →AD510LH/+
✅ Drop-InConsulte o preço
AD515AKH Classificações máximas e características elétricas
| Product Type | Monolithic precision, low-power FET-input electrometer operational amplifier |
| Maximum Input Bias Current | 150 fA |
| Typical Quiescent Current | 0.6 mA |
| Maximum Quiescent Current | 1.5 mA |
| Offset Voltage | 1.0 mV maximum |
| Number of Circuits | 1 circuit |
| Input Transistor Type | FET |
| Encapsulamento | TO-99-8 metal can |
| Número de pinos | 8 pins |
| Terminal Form | Wire |
| Temperature Grade | Commercial |
Guia de Configuração de Pinos do AD515AKH to-99-8 metal can
Configuração de pinos para AD515AKH (pacote to-99-8 metal can). Este componente analógico possui pinos de entrada, saída e alimentação. Consulte o datasheet do fabricante para configurações de offset null, compensação e pino de enable. Ideal para circuitos de condicionamento de sinal e amplificadores.
Nenhum dado detalhado de pinagem disponível para AD515AKH.
Consulte a ficha técnica para a configuração completa dos pinos.
Aplicações típicas
AD515AKH é adequado para 6 aplicaçãos: Electrochemical Sensor Interface, Photodiode Optical Measurement, Battery-Powered Precision Instrument, Chemical and Gas Sensor Monitoring, Capacitance and Charge Sensing, High-Impedance Voltage Divider.
Electrochemical Sensor Interface
The AD515AKH is well suited to pH electrodes, ion-selective electrodes, and other electrochemical sensors whose effective source impedance can be extremely high. Its 150 fA maximum input bias current limits the current drawn from the sensor, while the 0.6 mA typical quiescent current supports compact battery-powered analyzers. Place the amplifier close to the sensor, surround the input with guarding, and use low-leakage board materials. Feedback resistors should be selected to balance input-generated Johnson noise against the bandwidth required by the electrode’s response time.
Recomendado
Photodiode Optical Measurement
AD515AKH can amplify the small photocurrent generated by a photodiode without imposing a large input-bias load on the source. The 150 fA maximum bias current is particularly useful for wide-area or reverse-biased detectors operating at low light levels, while the maximum 1.5 mA supply current keeps instrumentation power manageable. Use the amplifier in a transimpedance configuration, keep the summing node compact, and provide guard rings around both the input and feedback resistor. Dark-current behavior, diode capacitance, and the required signal bandwidth determine the practical feedback resistance.
Recomendado
Battery-Powered Precision Instrument
The AD515AKH supports portable measurement systems because its typical quiescent current is 0.6 mA and its maximum is 1.5 mA. A single low-power FET-input amplifier can monitor battery current, high-resistance sensors, or environmental probes without consuming the power associated with a general-purpose precision amplifier. Use a low-noise regulator, maintain consistent grounding between the sensor and analog return, and include a power-on reset strategy if the measurement must start from a defined state. Battery chemistry, standby leakage, and cable leakage can ultimately dominate the complete instrument budget.
Recomendado
Chemical and Gas Sensor Monitoring
AD515AKH is appropriate for amperometric gas sensors, humidity elements, conductivity probes, and other transducers that produce signals through very high source impedances. Its 150 fA maximum input bias current reduces unintended interaction with the sensor, and the 1.5 mA maximum supply current supports compact field instruments. The design should include guarded input routing, contamination-resistant board assembly, and defined electrode-bias polarity. Filter capacitors must be selected to avoid excessive settling time or excessive leakage across the sensor itself.
Recomendado
Capacitance and Charge Sensing
AD515AKH can serve as the input amplifier for capacitive touch sensors, charge-sensitive detectors, and high-impedance capacitance-measurement systems. The low input bias current limits state-space charging of the sensing node, helping preserve charge between conversion cycles. Use short guarded traces, a defined virtual ground at the non-inverting input, and a feedback network whose resistance balances capacitor leakage against the required noise performance. The TO-99 package is mechanically robust for laboratory and industrial instrumentation, although the complete assembly must remain clean and dry.
Recomendado
High-Impedance Voltage Divider
AD515AKH is useful when measuring the midpoint of a very high-value resistive divider where the amplifier input current could otherwise create a significant measurement error. With maximum bias current of 150 fA, the device can present a very light load to high-impedance dividers while providing the accuracy benefits of laser-trimmed offset. Layout remains critical: use matched guarding, avoid flux residue around the input node, and select resistor tolerance according to the divider’s real accuracy requirement. Thermal effects in the high-value resistors may exceed the amplifier’s offset contribution.
Recomendado
Resumo dos produtos recomendados
Dados de referência de engenharia para AD515AKH — comparação, orientação de projeto e informações de conformidade.
Guia de seleção
Comparação com alternativas
| Parâmetro | Este produto | AD515ALH | AD515AJH | AD517SH | AD510LH |
|---|---|---|---|---|---|
| Brand | Analog Devices | Analog Devices | Analog Devices | Analog Devices | Analog Devices |
| Package | TO-99-8 metal can | TO-99-8 metal can | TO-99-8 metal can | TO-99-8 metal can | TO-99-8 metal can |
| Number of Pins | 8 pins | 8 pins | 8 pins | 8 pins | 8 pins |
| Amplifier Type | J-FET electrometer operational amplifier | AD515A-family FET-input operational amplifier | AD515A-family FET-input operational amplifier | J-FET operational amplifier | General-purpose operational amplifier |
| Number of Circuits | 1 circuit | 1 circuit | 1 circuit | 1 circuit | 1 circuit |
Principais diferenciais
- Lower guaranteed input bias current within the AD515A family (vs AD515AJH)
- Tighter leakage limit than the broader J-grade option (vs AD510LH)
- Lower power ceiling than AD517SH is not established (vs AD517SH)
Notas de projeto
Guard the high-impedance input and keep the sensitive node physically small. Route a guard structure driven by the amplifier’s non-inverting input potential around the input trace and feedback resistor where the application permits it. This reduces surface leakage and capacitive pickup that can exceed the AD515AKH’s 150 fA maximum bias current. Avoid placing high-impedance routes beside switching nodes, relay contacts, or mechanically dirty board areas.
Do not assume that a 150 fA input-current specification eliminates contamination effects. Flux residue, moisture, fingerprints, and connector leakage can create currents far above the amplifier bias current. Clean the assembled board using a process compatible with the sensor and include conformal coating when the operating environment is humid or dusty. Use a defined input bias path so the input cannot float during power-up, testing, or cable disconnection.
Estimated: at a 18 V supply and 1.5 mA maximum quiescent current, the amplifier quiescent dissipation is 27 mW, calculated as 18 V × 1.5 mA. Use a low-noise regulator and place its bypass capacitors near the supply pins. Stable reference and ground planes matter because supply variation and ground movement can appear as input-referred error in a high-gain electrometer measurement, even when the amplifier itself consumes little power.
Informações de conformidade
The manufacturer product description identifies a commercial-grade precision operational amplifier but does not provide RoHS, REACH, lead-free, halogen-free, conflict-minerals, or automotive qualification statements.