AS7331-AQFM 개요
A spectral UV sensor is an optical semiconductor that converts ultraviolet irradiance into a digital value. It sits within the sensor hierarchy: photodiode -> optical/ambient-light sensor -> spectral sensor -> environmental sensing subsystem, serving UV-index meters, disinfection monitors and laboratory instruments.
Key features include a high dynamic range adjustable up to 3.43E+10 via a 16-to-24-bit ADC, sensitivity up to 421 counts/(uW/cm2) in UVA, and a smallest LSB of 2.38 nW/cm2 at a 64 ms integration time. Irradiance responsivity is tunable through gain, conversion time and internal clock frequency, letting designers trade full-scale range against resolution. An integrated temperature sensor supports coarse compensation of thermal drift, and the I2C address is adjustable for multi-sensor buses.
Architecturally, the AS7331 converts photodiode currents through a low-noise analog front end into digital results, supporting continuous and triggered measurement modes with four low-power operating modes. The SYNC input enables externally synchronized sampling, while the RDY output (push-pull or open-drain) signals conversion completion - useful for deterministic sampling in process control.
Typical applications include UV index and personal exposure meters, UVC disinfection process monitoring, fluorescence detection, and UV measurement in mobile and wearable devices.
Design-wise, the glass or aperture area above the photodiodes must remain unobstructed, and the I2C bus should use appropriate pull-ups; the adjustable gain settings should be chosen so strong UVA sources do not saturate the ADC.
This page adds distributor pricing (as of 2026-10-11), drop-in alternative analysis, and design guidance beyond the manufacturer datasheet.
드롭인 대체품: AS7331-AQFM — 동일 패키지, 핀 호환. PCB 재작업이 필요한 다른 패키지 부품은 제외됩니다.
이 시리즈의 변형
다음과 드롭인 호환되는 동일 시리즈 모델: AS7331-AQFM (동일 폼 팩터 및 풋프린트) — 차이점: Interface, Operating Modes, Package, Sensor Type, Supply Voltage.
빠른 비교 도구 — 나란히 비교할 대체 부품을 선택하십시오:
이 제품에는 사용 가능한 드롭인 대체품이 없습니다.
대체품 요청AS7331-AQFM 최대 정격 & 전기적 특성
| 센서 유형 | 3-channel spectral UV sensor (UVA/UVB/UVC) |
| Spectral Range | 260 nm to 360 nm |
| 채널 수 | 3 (UVA, UVB, UVC) with interference filter technology |
| 인터페이스 | I2C (adjustable slave address) |
| 공급 전압 | 3.3 V |
| ADC Resolution | 16 to 24 bit |
| Dynamic Range | up to 3.43E+10 (adjustable) |
| Sensitivity (UVA) | up to 421 counts/(uW/cm2) |
| Smallest LSB | 2.38 nW/cm2 (at 64 ms integration time) |
| Operating Modes | 4 low-power modes; continuous and triggered measurement |
| Integrated Temperature Sensor | Yes (coarse thermal compensation) |
| Synchronization | SYNC input; RDY output (push-pull or open-drain) |
| 패키지 | 16-pin OLGA / TFLGA, SMD |
| 실장 유형 | Surface Mount |
| Packaging | Tape & Reel (LF T&R) |
AS7331-AQFM 16-pin olga / tflga, smd 핀 구성 가이드
AS7331-AQFM(16-pin olga / tflga, smd 패키지)의 핀 구성입니다. 핀 번호, 기능, 연결 다이어그램은 제조사 데이터시트에 정의되어 있습니다. 정확한 풋프린트와 솔더링 지침은 해당 문서를 참조하십시오.
다음에 대한 상세 핀아웃 데이터 없음: AS7331-AQFM.
다음을 참조하십시오: 데이터시트 전체 핀 구성은 다음을 참조하십시오.
일반적인 애플리케이션
AS7331-AQFM 적합 용도: 6 애플리케이션s: UV Index and Personal Exposure Meters, UVC Disinfection Process Monitoring, Fluorescence Detection, Mobile Device UV Measurement, Laboratory and Industrial UV Instrumentation, Smart Home and IoT UV Monitoring.
UV Index and Personal Exposure Meters
The AS7331-AQFM fits solar UV exposure measurement because its three filtered channels (260-360 nm) separate UVA and UVB the way UV index weighting requires, with UVB isolation provided by on-chip interference filters rather than external optics. Its sensitivity up to 421 counts/(uW/cm2) in UVA and smallest LSB of 2.38 nW/cm2 (64 ms integration) let a wearable resolve faint diffuse daylight, while the adjustable 16-to-24-bit conversion chain prevents saturation under direct midday sun. In a wearable or handheld dosimeter, the sensor is polled over I2C in continuous mode at low duty cycle from one of its four low-power operating modes to preserve battery life, and the integrated temperature sensor provides coarse compensation of responsivity drift across outdoor temperature swings.
추천
UVC Disinfection Process Monitoring
UVC germicidal systems need closed-loop verification that 260-280 nm dose is actually delivered, and the AS7331-AQFM's dedicated UVC channel with interference-filter isolation directly answers this. Its high dynamic range (up to 3.43E+10) lets one sensor design cover both close-range high-irradiance fixtures and distant low-irradiance verification points by reprogramming gain and conversion time. The SYNC input allows triggered sampling synchronized to lamp PWM, and the RDY output (push-pull or open-drain) gives the process controller deterministic conversion-complete timing. In a disinfection cabinet, the sensor is mounted where the load sits, read over I2C after each SYNC pulse, and dose is integrated in firmware from the nW/cm2-scale LSB resolution, enabling per-cycle pass/fail logging.
추천
Fluorescence Detection
Fluorescence assays emit weak light against a dark background, and the AS7331-AQFM's 2.38 nW/cm2 smallest LSB at 64 ms integration gives it the sensitivity to resolve these faint signals in the UVA/UVB bands where many fluorophores are excited. Because gain, conversion time and internal clock frequency are independently programmable, the analog front end can be tuned per assay: high gain with long integration for low-concentration samples, reduced gain for concentrated ones, all without changing hardware. The triggered measurement mode lets the system fire the excitation source, then sample only during the emission window using the SYNC input, suppressing ambient background. The I2C interface with adjustable address supports multi-well or multi-channel reader architectures on one bus.
추천
Mobile Device UV Measurement
Smartphones and tablets increasingly expose UV sensing for weather and skin-health apps, and the AS7331-AQFM targets exactly this use case per the ams OSRAM product page. Its compact 16-pin OLGA SMD package fits the tight bezel and camera-island footprints of mobile hardware, while four low-power operating modes limit average current draw in always-listening exposure tracking. The I2C interface with adjustable address coexists with other phone sensors on the shared bus, and the 3.3 V supply matches common mobile rails. Integration effort is minimal: no external filters or optics are needed because each channel's interference filter is built into the die, so a smartphone maker gets calibrated three-band UV data from a single surface-mount part.
추천
Laboratory and Industrial UV Instrumentation
Bench and inline instruments benefit from the AS7331-AQFM's quantifiable, programmable measurement chain: a 16-to-24-bit conversion with dynamic range up to 3.43E+10 and configurable LSB down to 2.38 nW/cm2 allows one design to span trace-level lamp leak checks and full-power irradiance calibration. In a laboratory dosimeter, the sensor sits behind a cosine-corrector, is triggered via SYNC for repeatability, and streams results over I2C to a logging MCU; the RDY flag enables gap-free acquisition. The integrated temperature sensor supports coarse thermal compensation between calibrations. Because channels are filtered on-die, no per-band optical alignment is required during manufacturing, which reduces production calibration cost compared to discrete-filter photodiode assemblies.
추천
Smart Home and IoT UV Monitoring
IoT nodes that report UV conditions - garden sensors, pool-side stations, smart window controllers - can use the AS7331-AQFM as a drop-in three-band UV front end. Its I2C output connects directly to common wireless MCU platforms, and the adjustable I2C address permits several sensor nodes on one controller bus. The four low-power modes suit battery-operated nodes: the sensor sleeps between periodic conversions and uses RDY to wake the host only when data is ready, cutting average power. With an LSB of 2.38 nW/cm2 the node resolves overcast-day UV, while the extended dynamic range keeps readings valid in direct sun. Cloud dashboards can then display UVA/UVB/UVC separately, which a single broad-band UV sensor cannot provide.
추천
추천 제품 요약
엔지니어링 참조 데이터: AS7331-AQFM — 비교, 설계 지침 및 규정 준수 정보.
선택 가이드
대체품과의 비교
| 파라미터 | 이 제품 |
|---|---|
| Package | 16-pin OLGA / TFLGA, SMD |
| Brand | ams OSRAM |
| Spectral Range | 260 nm to 360 nm (UVA/UVB/UVC) |
| Channels | 3 (UVA, UVB, UVC), interference-filtered |
| ADC Resolution | 16 to 24 bit |
| Dynamic Range | up to 3.43E+10 |
| Interface | I2C, adjustable address |
| Sensitivity (UVA) | up to 421 counts/(uW/cm2) |
| Smallest LSB | 2.38 nW/cm2 at 64 ms integration |
| Unit Price (qty 1, as of 2026-10-11) | $9.69 (LCSC) |
주요 차별점
- True 3-band spectral separation on-die (vs LTR390-UV)
- Extreme dynamic range with programmable resolution (vs discrete UV photodiode + ADC)
- Trade-off: single-sourced niche package (vs LTR390-UV)
설계 노트
Place the AS7331-AQFM so its optical aperture has an unobstructed field of view, and keep the OLGA-16 land pattern per the ams OSRAM datasheet footprint recommendation. Because the interference filters are on-die, no external filter glass is needed, but avoid placing UV-absorbing cover materials (polycarbonate blocks UVC strongly; use quartz or UV-transmitting acrylic if a window is required). Keep the sensor away from board edges where solder wicking could contaminate the window during reflow.
Use appropriate I2C pull-up resistors (typically 2.2k to 10k ohm depending on bus capacitance) and keep the bus short. Take advantage of the adjustable I2C address when networking multiple AS7331 sensors. Route the RDY line as a dedicated interrupt to the host MCU and configure it open-drain if it shares an interrupt line with other devices; use push-pull for lowest latency in deterministic triggered-sampling systems driven from SYNC.
Choose gain and conversion time so expected maximum irradiance does not saturate the ADC: direct sunlight UVA can exceed 1000 uW/cm2, so verify headroom against the 421 counts/(uW/cm2) maximum sensitivity setting. Do not rely on the integrated temperature sensor for precision correction - it is specified for coarse compensation only, so perform system-level calibration for demanding photometric use. Confirm the exact supply voltage and operating temperature range from the datasheet before finalizing power-tree design.
규정 준수 정보
DigiKey listing shows 'OLGA16 LF T&R' suffix; LF is typically the lead-free designation but explicit RoHS/REACH compliance statements were not present in the verified data - confirm on the ams OSRAM product page.