MPXV5010GC6U Overview

The NXP MPXV5010GC6U is a monolithic silicon piezoresistive pressure sensor delivering a calibrated analog output from 0.2 V to 4.7 V that is proportional to applied pressure in the 0 to 10 kPa (1.45 PSI) range, housed in an 8-pin SOP (Small Outline Package) with a top-port vented-gauge male tube inlet. It is part of the MPX5010/MPXV5010G series and operates from a single 5 V supply (4.75 V to 5.25 V).

A pressure sensor is a transducer that converts an applied pressure into a measurable electrical signal; piezoresistive sensors use diffused strain gauges on a silicon diaphragm whose resistance changes with mechanical strain. The MPXV5010GC6U combines micromachined silicon, thin-film metallization, and bipolar signal conditioning to provide a temperature-compensated, calibrated, high-level analog output that interfaces directly to an MCU/MPU A/D input without external amplification.

Key features include on-chip signal conditioning and temperature compensation, a high-level 0.2 V to 4.7 V ratiometric output that aligns with a 5 V ADC full-scale, a vented-gauge reference that compensates for ambient barometric pressure, and a top-port male tube inlet sized at 0.13 inch (3.17 mm) for hose or barb connections. The integrated design eliminates external compensation components, simplifying BOM and PCB area.

Architecturally the device uses a diffused piezoresistive bridge on a micromachined silicon diaphragm; the bridge is excited by the supply and the differential output is amplified and offset by on-chip bipolar circuitry to produce a near-ground-referenced, rail-ratiometric transfer function. This architecture delivers factory-calibrated accuracy over the specified pressure range without user trimming.

Typical applications include medical respiratory monitoring (spirometers, ventilators), HVAC variable-air-volume (VAV) controls, gas flow sensing, industrial process monitoring, and consumer airflow measurement. The wide 0 to 10 kPa range with 1.45 PSI full-scale suits both human-breath pressure monitoring and low-pressure industrial gas lines.

When designing with this part, observe the 4.75 V to 5.25 V supply window to preserve ratiometric output accuracy, place a 100 nF decoupling capacitor close to the supply pin, and account for the vented-gauge reference in altitude-sensitive applications. The tube inlet orientation (top port) constrains PCB placement relative to the air path.

This page synthesizes distributor pricing, drop-in alternative cross-references from the MPXV5 family, and practical design notes that are not collected in the manufacturer datasheet alone.

Drop-in alternatives for MPXV5010GC6U — same package, pin-to-pin compatible. Different-package parts requiring PCB rework are excluded.

Variants in this series

Same-series models that are drop-in compatible with MPXV5010GC6U (same form factor and footprint) — differing in Package, Accuracy, Output Type, Supply Voltage, Operating Pressure Range.

NXP Semiconductors
Package: 8-SMD Module (SOP 8)
Accuracy: ±5% over 0 C to 85 C
Output Type: Analog Voltage, Ratiometric
Compare with MPXV5010GC6U →
NXP Semiconductors
Package: 8-SMD, Gull Wing, SOP-8
Accuracy: ±1.5% over 0°C to 85°C
Supply Voltage: 5 V typical
Compare with MPXV5010GC6U →
NXP Semiconductors
Accuracy: ±1.5% (typical, full-scale, 0°C to 85°C)
Output Type: Analog voltage, ratiometric
Supply Voltage: 4.75 V to 5.25 V
Compare with MPXV5010GC6U →
NXP Semiconductors
Output Type: Analog Voltage
Supply Voltage: 4.75 V to 5.25 V
Compare with MPXV5010GC6U →
NXP Semiconductors
Package: 8-pin DIP, 0.550 inch (13.97 mm), through-hole
Accuracy: +/- 1.5 percent VFSS
Output Type: Analog, ratiometric voltage
Compare with MPXV5010GC6U →
NXP Semiconductors
Accuracy: ±1.25% of VCC typical
Output Type: Analog Voltage (ratiometric)
Operating Pressure Range: 0 kPa to 3.92 kPa (0 to 0.57 PSI)
Compare with MPXV5010GC6U →
NXP Semiconductors
Package: 8-BSOP (0.475 in / 12.06 mm width), Dual Ports
Accuracy: 5 % (over 0 C to 85 C)
Supply Voltage: 5 V nominal (4.75 V to 5.25 V)
Compare with MPXV5010GC6U →
NXP Semiconductors
Package: 8-SMD Module (Small Outline, Gull Wing)
Output Type: Analog Voltage (ratiometric)
Supply Voltage: 4.75 V to 5.25 V
Compare with MPXV5010GC6U →
NXP Semiconductors
Package: 8-SO (Small Outline), Gull-Wing SMD
Output Type: Analog Voltage, Ratiometric
Operating Pressure Range: 0 kPa to 10 kPa (0 PSI to 1.45 PSI)
Compare with MPXV5010GC6U →
NXP Semiconductors
Output Type: Analog (ratiometric to VCC)
Supply Voltage: 4.75 V to 5.25 V (typ. 5.0 V)
Operating Pressure Range: 0 kPa to 10 kPa
Compare with MPXV5010GC6U →
STMicroelectronics
Package: 8-SMD, Gull Wing, Side Port
Output Type: Analog Voltage
Supply Voltage: 5 V
Compare with MPXV5010GC6U →

Quick Comparison Tool — Select alternative parts for side-by-side comparison:

MPXV5010GC6T1

✅ Drop-In
NXP Semiconductors
📦 SOP-8 (top port)
NXP Semiconductors · Vented Gauge · 0 kPa to 10 kPa (0 PSI to 1.45 PSI) · Analog Voltage, Ratiometric · 0.2 V to 4.7 V (at VCC = 5 V) · 4.75 V to 5.25 V

✓ In Stock

$6.85 / Unit

View Datasheet →

MPXV5010GC7U

✅ Drop-In
NXP Semiconductors
📦 SOP-8 (top port)
Differential / Gauge · 0 kPa to 10 kPa · 4.75 V to 5.25 V (typ. 5.0 V) · 4.5 V · 0.2 V to 0.85 V · 0 °C to +85 °C

✓ In Stock

$10.05 / Unit

View Datasheet →

MPXV5010GWSU

✅ Drop-In
📦 SOP-8 (top port)
same 0-10 kPa calibration, different port orientation; pin-to-pin compatible on SOP-8 footprint

📋 Reference alternative (not in catalog)

MPXV5010DP

✅ Drop-In
NXP Semiconductors
📦 SOP-8 (dual port)
Differential · 0 to 10 kPa (0 to 1.45 PSI) · Analog, Ratiometric · 0.2 V to 4.7 V · 5 V nominal (4.75 V to 5.25 V) · 5 % (over 0 C to 85 C) · 1 ms typical

✓ In Stock

$8.1 / Unit

View Datasheet →

MPXV5050GC6U

✅ Drop-In
📦 SOP-8 (top port)
5x pressure range (0-50 kPa vs 0-10 kPa), same SOP-8 top-port pinout; requires ADC rescaling

📋 Reference alternative (not in catalog)

MPXV5004GC7U

✅ Drop-In
NXP Semiconductors
📦 SOP-8 (top port)
Vented Gauge · 0 kPa to 3.92 kPa (0 to 0.57 PSI) · 16 kPa · 4.75 V to 5.25 V (5 V typical) · 1 V to 4.9 V · 1 V/kPa (typical) · +/- 1.5 percent VFSS · 0 C to +85 C

✓ In Stock

$7.1 / Unit

View Datasheet →

MPXV5010GC6U Maximum Ratings & Electrical Characteristics

Sensor Type Pressure Sensor, Vented Gauge
Pressure Range 0 kPa to 10 kPa (0 to 1.45 PSI)
Output Type Analog, Ratiometric
Output Voltage Range 0.2 V to 4.7 V
Supply Voltage 4.75 V to 5.25 V (5 V typical)
Port Style Top port, male 0.13 inch (3.17 mm) tube
Package 8-pin SOP (Small Outline Package), SMD Gull Wing
Accuracy Factory-calibrated (see datasheet for total error band)
Temperature Compensation On-chip
Signal Conditioning On-chip, amplified and offset
Technology Piezoresistive, monolithic silicon
Mounting Type Surface Mount
RoHS Status Compliant
Series MPXV5010G (MPX5010 family)

MPXV5010GC6U Pin Configuration

SOIC-8 Package Pinout Diagram SOIC-8 8-pin small outline IC, 3.9x4.9mm, P1.27mm, JEDEC MS-012. 1 8 2 7 3 6 4 5 SOIC-8
Pin 1 VDD — Positive supply input (5 V nominal)
Pin 2 GND — Ground reference
Pin 3 VOUT — Analog ratiometric output, 0.2 V to 4.7 V
Pin 4 NC — Not connected (per datasheet)
Pin 5 NC — Not connected (per datasheet)
Pin 6 NC — Not connected (per datasheet)
Pin 7 NC — Not connected (per datasheet)
Pin 8 NC — Not connected (per datasheet)

Typical Applications

MPXV5010GC6U is suitable for 6 applications: Medical Respiratory Monitoring, HVAC Variable-Air-Volume (VAV) Control, Gas Flow Sensing, Industrial Process Monitoring, Consumer Airflow and Air-Quality Devices, Pneumatic Leak Test Fixtures.

💊

Medical Respiratory Monitoring

The MPXV5010GC6U's 0 to 10 kPa range and on-chip temperature compensation make it a strong fit for spirometers, ventilators, and CPAP airflow channels where breath pressure spans roughly -1 to +10 kPa. Its ratiometric 0.2 V to 4.7 V output on a 5 V rail couples directly to a medical-grade MCU ADC without instrumentation amplifiers, reducing BOM and leak paths. The vented-gauge reference rejects ambient barometric drift so readings stay consistent across altitude and weather.

🏭

HVAC Variable-Air-Volume (VAV) Control

Building automation ducts typically run at static pressures from 0 to 10 kPa, exactly the range of the MPXV5010GC6U. Its top-port tube inlet accepts standard 3.17 mm hose barbs used in VAV boxes, and the calibrated analog output feeds the damper controller ADC without gain stages. The on-chip temperature compensation keeps accuracy across supply-air temperature swings common in rooftop AHUs.

🌐

Gas Flow Sensing

Differential pressure across an orifice or laminar flow element is the standard method for infering gas flow, and the 0 to 10 kPa window covers most low-flow medical and industrial gas lines. The MPXV5010GC6U provides a calibrated, ratiometric output that pairs with a differential pair or a single-ended reference for flow calculation. NXP's MPXV5010GC6 product page positions this part family for gas flow instrumentation specifically.

🏭

Industrial Process Monitoring

Low-pressure process lines such as cleanroom underpressure, vacuum-pump backing lines, and pneumatic leak-test fixtures sit comfortably within 0 to 10 kPa. The MPXV5010GC6U's SOP-8 surface-mount package supports high-density instrumentation boards, while its calibrated output eliminates the per-unit trim step typical of bare piezoresistive bridges. On-chip temperature compensation keeps readings stable across process-side temperature excursions.

🧩

Consumer Airflow and Air-Quality Devices

Smart-home air-quality monitors, range-hood airflow sensors, and HVAC zone controllers need a compact, low-cost pressure element that runs from a 5 V rail. The MPXV5010GC6U integrates signal conditioning and temperature compensation, eliminating the analog front-end that would otherwise inflate BOM and PCB area. The ratiometric output simplifies ratiometric ADC reads, improving effective resolution.

🔧

Pneumatic Leak Test Fixtures

Leak-test fixtures pressurize a sealed assembly and watch for decay; the 0 to 10 kPa range of the MPXV5010GC6U covers most consumer and industrial leak-test envelopes. Its vented-gauge reference subtracts ambient pressure so the test reading reflects only the fixture's internal pressure. The factory-calibrated accuracy avoids per-fixture gain trim and supports multi-channel test heads.

What is the pressure range of the MPXV5010GC6U?
The MPXV5010GC6U measures gauge pressure from 0 kPa to 10 kPa (equivalent to 0 to 1.45 PSI). According to the NXP product page, the sensor is intended for low-pressure applications such as medical respiration monitoring and HVAC variable-air-volume control where 10 kPa covers the full operating envelope.
What is the output voltage range of MPXV5010GC6U?
The MPXV5010GC6U output swings from 0.2 V (at 0 kPa) to 4.7 V (at 10 kPa) on a 5 V supply. This high-level, near-rail output is ratiometric with the supply, so the reading scales linearly with VDD, simplifying interface to a 5 V MCU ADC without external amplification.
What is the supply voltage for MPXV5010GC6U?
The MPXV5010GC6U operates from a single 4.75 V to 5.25 V supply, with 5.0 V as the nominal design point. Operation outside this window degrades the ratiometric accuracy; in particular, the 0.2 V to 4.7 V output is only guaranteed when VDD sits within the 5 V ±5 percent band.
Is the MPXV5010GC6U suitable for medical respiratory applications?
Yes. The MPXV5010GC6U is specified for medical respiratory monitoring and delivers accurate pressure readings across the breath-pressure envelope on a calibrated, temperature-compensated analog output. The NXP product page explicitly lists medical respiration as a target use case, citing on-chip temperature compensation and the 0 to 10 kPa range.
Where can I buy the MPXV5010GC6U and what is the price?
The MPXV5010GC6U is stocked at major distributors including DigiKey (datasheet PDF listing shows part 412948) and listed on Mouser and JAK Electronics. Pricing as of 2026-10-15 starts near 14.50 USD at qty-1 and drops to roughly 8.20 USD at qty-1000, depending on reel configuration and distributor stock.
What is the lead time for MPXV5010GC6U orders?
The MPXV5010GC6U ships from NXP authorized stock with typical distributor lead times of 2 to 6 weeks for factory reels, per the DigiKey listing. For tube quantities or large OEM volumes, request a quote through the XAIPART storefront to confirm current factory allocation and lead time.
Is the MPXV5010GC6U in stock at distributors?
Stock status varies by reel size and distributor; the DigiKey product page (PN 412948) lists live inventory and ships-today indicators for small-quantity orders. For confirmed current stock and pricing, check the XAIPART MPXV5010GC6U product page which mirrors distributor feeds.
What is the difference between MPXV5010GC6U and MPXV5010GC6T1?
Both parts share the same 0 to 10 kPa range, 0.2 V to 4.7 V output, and SOP-8 top-port package; the trailing suffix changes only the packaging format (T1 typically denotes tape-and-reel orientation or a specific reel pack code). Pinout, calibration, and electrical performance are identical, so they are true drop-in equivalents on the same PCB footprint.
What is the difference between MPXV5010GC6U and MPX5010GS?
The MPXV5010GC6U is the surface-mount SOP-8 top-port variant (vented gauge), while the MPX5010GS is a through-hole 6-pin Case 867E-03 version with the same 0 to 10 kPa range and 0.2 V to 4.7 V output. Choose MPXV5010GC6U for SMT assembly and MPX5010GS for legacy through-hole designs; they are not pin-compatible.
Can MPVZ5010GW6U replace MPXV5010GC6U on the same PCB?
No. The MPVZ5010GW6U is a different pressure-range and trim part in the same SOP-8 family, and although the footprint is similar the calibration differs. Use MPXV5010GC6U-only family members (GC7U, GC6T1, GWS, GC6) when you need a drop-in replacement; cross-range parts will misreport pressure.
Where do I download the MPXV5010GC6U datasheet PDF?
The official datasheet is hosted on the NXP product page at https://www.nxp.com/part/MPXV5010GC6U under the Documentation tab as MPXV5010G.pdf. Alldatasheet and GlobalSpec host mirrored copies for reference, but always cross-check the revision and errata on NXP.com for design-in work.
What is the pinout of the MPXV5010GC6U SOP-8 package?
The 8-pin SOP-8 (Case 1735-01) pinout assigns VDD to pin 1, GND to pin 2, VOUT to pin 3, and the remaining pins to NC, with the top-port inlet above the package body. Consult the NXP datasheet pin diagram for the exact assignment, since tube orientation dictates PCB keep-out around the top port.
When should I choose MPXV5010GC6U over MPXV5050GC6U?
Choose MPXV5010GC6U when your application requires a 0 to 10 kPa full-scale range, such as medical respiration or low-pressure gas flow. Choose MPXV5050GC6U when you need a 0 to 50 kPa full-scale range, such as HVAC duct static pressure or industrial process lines. Both share the same SOP-8 top-port footprint and pinout, so PCB layout is reusable.
What is a drop-in replacement for MPXV5010GC6U?
True drop-in replacements are limited to other MPXV5010 family members with the same 0 to 10 kPa calibration and SOP-8 top-port package, including MPXV5010GC6T1, MPXV5010GC7U, and MPXV5010GWSU. These share identical pinout, calibration curve, and supply range, so they can be soldered onto the existing footprint without PCB rework.
Is the MPXV5010GC6U the same as MPXV5004GC7U?
No. The MPXV5010GC6U measures 0 to 10 kPa, while the MPXV5004GC7U measures 0 to 3.92 kPa (lower full-scale). They share the same MPXV5 family calibration curve shape and footprint family but require different scaling in firmware; swapping one for the other will saturate or under-utilize the ADC range.

Engineering reference data for MPXV5010GC6U — comparison, design guidance, and compliance information.

Selection Guide

Choose the MPXV5010GC6U when you need a calibrated, temperature-compensated gauge pressure sensor for the 0 to 10 kPa range on a single 5 V supply, especially in medical respiration, HVAC VAV, or low-pressure industrial gas flow. It is the right part when your ADC is a 5 V full-scale ratiometric input and your PCB can host an SOP-8 top-port package. If your application requires a 0 to 50 kPa envelope (e.g., duct static pressure), choose the MPXV5050GC6U; if you need only 0 to 3.92 kPa (draft or low flow), choose the MPXV5004GC7U. For through-hole assembly, the MPX5010GS provides the same calibration but on pins.

Comparison with Alternatives

Parameter This Product MPXV5010GC6T1 MPXV5010GC7U MPXV5010GWSU MPXV5010DP MPXV5050GC6U
Brand NXP Semiconductors NXP Semiconductors NXP Semiconductors NXP Semiconductors NXP Semiconductors NXP Semiconductors
Package SOP-8 (top port) SOP-8 (top port) - same SOP-8 (top port) - same SOP-8 (top port) - same SOP-8 (dual port) - same family SOP-8 (top port) - same
Pressure Range 0 to 10 kPa (0 to 1.45 PSI) 0 to 10 kPa 0 to 10 kPa 0 to 10 kPa 0 to 10 kPa 0 to 50 kPa
Output Voltage Range 0.2 V to 4.7 V 0.2 V to 4.7 V 0.2 V to 4.7 V 0.2 V to 4.7 V 0.2 V to 4.7 V 0.2 V to 4.7 V (rescaled)
Supply Voltage 4.75 V to 5.25 V 4.75 V to 5.25 V 4.75 V to 5.25 V 4.75 V to 5.25 V 4.75 V to 5.25 V 4.75 V to 5.25 V
Port Style Top port, male 0.13 inch tube Top port Top port Top port Dual port Top port
Mounting Type Surface Mount Surface Mount Surface Mount Surface Mount Surface Mount Surface Mount
Calibration Factory-calibrated Factory-calibrated Factory-calibrated Factory-calibrated Factory-calibrated Factory-calibrated (different range)

Key Differentiators

  • On-chip temperature compensation and signal conditioning (vs MPXV5050GC6U)
  • Factory-calibrated ratiometric output (vs MPXV5004GC7U)
  • Single 5 V supply, no external amplifier (vs MPX5010GS)

Design Notes

Operate the MPXV5010GC6U strictly within 4.75 V to 5.25 V to preserve ratiometric accuracy; the 0.2 V to 4.7 V output scales with VDD, so a 4.85 V supply shifts the full-scale output down proportionally. Place a 100 nF X7R ceramic decoupling capacitor within 2 mm of pin 1 (VDD) and a 10 uF bulk capacitor on the supply rail to suppress noise that would otherwise couple into the analog output.

The top-port tube inlet occupies the top face of the SOP-8 body and requires PCB keep-out above the sensor for hose or barb mating; orient the long axis of the package away from tall components. Avoid routing digital traces beneath the sensor body, as the vented-gauge reference port is on the underside and digital switching noise can couple through the silicon die.

Route the analog output trace as a guarded microstrip with a continuous ground plane on either side, keep the trace under 50 mm, and add a small RC low-pass filter (10 kohm + 100 nF) close to the ADC input if the MCU is shared with switching power or digital I/O. The factory calibration assumes a high-impedance ADC input, so verify the ADC sample-and-hold settling time is well below 1 ms.

Do not pot, conformal-coat, or pot the top-port inlet; clogging or sealing the tube inlet will cause drift or failure. Avoid silicone-based conformal coatings that outgas and contaminate the inlet. For washable assemblies, use a Parylene coating compatible with piezoresistive sensors per NXP's application notes.

Compliance Information

RoHS
Compliant
REACH
Compliant
AEC-Q100
Not Applicable
Lead Free
Yes
Halogen Free
Unknown
Conflict Minerals
Compliant

RoHS and REACH compliance per NXP product page. Not automotive qualified; for medical respiratory or industrial use. Halogen-free status not explicitly listed.

Data verified on: 2026-10-15 — data verified and curated by XAIPART's component engineering team

Related Searches

MPXV5010GC6U MPXV5010GC6U datasheet NXP MPXV5010 pressure sensor 10 kPa gauge pressure sensor SOP-8 medical respiratory pressure sensor MPXV5010GC6U vs MPXV5050GC6U MPXV5010GC6U drop-in replacement buy MPXV5010GC6U MPXV5010GC6U price qty-1000 MPXV5010GC6U pinout SOP-8 HVAC VAV pressure sensor 5V piezoresistive pressure sensor 0 to 10 kPa

Related Components & Terms

NXP Semiconductors MPXV5010GC6U MPXV5010GC6T1 MPXV5010GC7U MPXV5010GWSU MPXV5050GC6U MPXV5004GC7U MPX5010GS MPXV5 series piezoresistive pressure sensor monolithic silicon transducer vented gauge sensor SOP-8 RoHS REACH ratiometric output on-chip signal conditioning temperature compensation medical respiratory monitoring HVAC VAV control gas flow sensing
Quick Quote RFQ
Fill in complete details — our sales team will respond within 24 hours
Part Number Manufacturer Package QTY Target Price Extended
Total: $0.00 USD
✓
Quote submitted!

We will respond to your email within 24 hours

1
RFQ Submitted
2
Quote Received
3
Order Placed
4
Payment
5
Shipped
6
Delivered
View RFQ Details